Blue_gradient

Purair® SAFESEARCH Package, Mail & Evidence Inspection Workstations

Controlled Airflow & Filtration for Safer Examination

Purair® SAFESEARCH provides a controlled, filtered workspace for the examination of mail, packages and evidence where hazardous drugs, powders, particulates or compatible chemical vapors may be encountered or leaking. 

Designed for government agencies, law enforcement, correctional facilities, forensic laboratories, mailrooms and secure receiving operations, SAFESEARCH combines inward containment airflow, activated-carbon filtration and HEPA particulate filtration in a self-contained workstation that requires no external ductwork.

Air is drawn away from the operator and into the enclosure at a nominal 100 fpm face velocity. The airstream passes through the Multiplex™ Filtration System, where selected activated carbon adsorbs compatible chemical vapors and HEPA filtration captures powders and particulates before filtered air is returned to the room.

24" W
36" W
48" W

Key Features

Why Choose the Purair SAFESEARCH?

Purair SAFESEARCH is a ductless workstation designed for routine screening and controlled examination of mail, packages and evidence. It combines inward airflow and filtration with a clear view of the work area and a high-contrast surface that makes light-colored powders easier to see.

Clear side and rear panels provide a broad view of the work area, while a hinged front sash gives operators access to packages and evidence and a black polypropylene tray makes light-colored powders easier to spot.

Air is drawn into the enclosure at a nominal face velocity of 100 fpm and passes through the Multiplex™ Filtration System, where HEPA filtration captures powders and particulates and selected activated carbon adsorbs compatible chemical vapors.

Available in 24, 36 and 48 inch widths, with standard-depth XT and reduced-depth XTS models, SAFESEARCH fits a range of countertops without requiring external exhaust ductwork.

IMPORTANT

Purair SAFESEARCH is intended for routine screening and controlled examination. An item already identified as a suspicious or potentially explosive, radiological, biological or chemical threat should not be opened or further manipulated. Follow the organization’s established emergency-response procedures.

When Should I Choose Purair SAFESEARCH?

For laboratories where space is limited.

Designed for controlled applications handling smaller volumes.

A compact option when the larger ADVANCED platform isn’t required.

When Should I Choose Purair® ADVANCED?

38″ internal working height for larger equipment and processes.

Designed for more demanding laboratory applications.

Seven standard sizes from 30″ to 96″ wide.

Compare Standard Ductless Fume Hoods

Ductless Fume Hoods
Available External Width
20.72 - 29.75"
24 - 48"
29.5 - 96"
24 - 48"
24 - 59.5"
Available External Height
15.67 - 21.16"
35"
53"
31.5"
78 - 87"
Available External Depth
12.02 - 24.72"
24 - 27"
27.375"
22.75"
27 - 28.5"
Available Internal Height
11 - 20.63"
23.6"
38"
15.75"
24 - 38"
Description
Fume Adsorber
Compact, Value
Premium, High Performance
Unique Downward Airflow
Educational, Demonstration

How Does a Purair SAFESEARCH Work?

Room air is drawn inward through the hood opening to help contain contaminants generated inside the enclosure. The contaminated air passes through an electrostatic pre-filter and the selected Multiplex™ Filtration System. Activated carbon adsorbs compatible chemical vapors, while optional HEPA or ULPA filtration captures particulates. The EC blower then returns filtered air to the laboratory.

How Does Multiplex™ Filtration Technology Work?

At the heart of Purair SAFESEARCH is the Multiplex™ Filtration System. It combines an electrostatic pre-filter, a primary filter and, where required, an optional secondary or stacked filter. The system can be configured for chemical adsorption, particulate filtration or applications requiring both.

The electrostatic pre-filter helps protect the main filter from dust, mists and particulates. Activated carbon filters adsorb compatible chemical vapors, while optional HEPA or ULPA filters capture fine particulates. Different filter media can be selected or combined based on the chemicals, operating conditions and filtration requirements of the application.

Multiplex Filtration Diagram
Ductless fume hood service

Application-Specific Filter Options

Purair SAFESEARCH can be configured with application-specific filtration for different laboratory needs. Activated carbon filters adsorb compatible chemical vapors, while optional HEPA/ULPA filtration captures powders and particulates. Filter selection depends on the chemicals and operating conditions of the application.

Filter Formula

Designed For

GP Plus!

The most widely used filter in the range, primarily for solvent, organic and alcohol removal.

ACI Plus!/SUL

Designed to neutralize volatile inorganic acid vapors.

ACR

Iodine and methyl iodide vapors

ACM

Mercury vapor

AMM

Removes vapors from dilute ammonia solutions and to remove low molecular weight amines.

FOR

Designed to oxidize formaldehyde and glutaraldehyde fumes; It is widely used in hospital pathology laboratories.

HEPA/UPLA

Powders and particulates.

Mail, Package & Evidence Inspection Applications

Purair SAFESEARCH provides a filtered workspace for routine mail and package screening and controlled evidence examination. Its ductless design fits into existing mailrooms, secure receiving areas and forensic facilities without external exhaust ductwork.

  • Government and institutional mailrooms
  • Secure corporate mailrooms
  • Correctional facility mail screening
  • Airport and transportation facility mail screening
  • Law enforcement evidence examination
  • Forensic laboratory evidence examination
  • Controlled examination of powders and residues
  • Vendor deliveries and secure receiving

Frequently Asked Questions

Below are answers to our frequently asked questions and links to additional helpful information.

A ductless package, mail and evidence inspection workstation designed to increase operator protection from hazardous drugs, compatible chemical vapors and particulates during controlled examination.

Air is drawn inward through the work opening and away from the operator. The airstream then passes through selected activated-carbon and HEPA filtration before filtered air is returned to the room.

No. SAFESEARCH is a self-contained ductless workstation.

Yes. HEPA filtration is used to capture powders and particulates and is rated 99.97% efficient at 0.3 microns.

No. Chemical vapors require appropriate activated-carbon filtration.

SAFESEARCH is designed for applications where hazardous drugs and particulates may be encountered during package examination. The specific substance, quantity and procedure should be reviewed before use, and appropriate PPE and work practices remain necessary.

Stop the examination and follow the facility’s unknown-substance or suspicious-mail response procedure. SAFESEARCH does not replace an incident-response plan.

No. Do not open or further manipulate an item already identified as a potential explosive, radiological, biological, chemical or other serious threat. Follow the organization’s emergency-response procedure.

The black polypropylene work tray provides greater visual contrast for white and light-colored powders.

Yes. An optional mobile base with locking casters is available.

Not Sure Which Ductless Fume Hood or Filter You Need?

Air Science can review your application, chemicals, concentrations, quantities, and process requirements to help determine an appropriate ductless fume hood and filtration configuration.