PCR Workstations &
UV Dead Air Boxes

Two Approaches to PCR, DNA/RNA &
Sample Contamination Control

Two Approaches to Sample Protection

Air Science offers two dedicated workstation platforms for contamination-sensitive PCR and molecular-biology workflows. Purair PCR Workstations use HEPA-filtered vertical airflow to create an ISO Class 5 clean work zone. Purair UV Dead Air Boxes use a circulation-free still-air enclosure with built-in 254 nm UV for surface decontamination between procedures.

Both are designed to help protect samples, reagents and processes from contamination in non-hazardous applications. The right choice depends on whether your workflow requires filtered clean airflow or a still-air UV enclosure.

IMPORTANT

Purair PCR Workstations and Purair UV Dead Air Boxes are product/sample-protection workstations, not personnel-containment devices. They should not be used as biological safety cabinets or chemical fume hoods when personnel protection from biohazards, hazardous aerosols, gases or chemical vapors is required.

How Does a PCR Workstation Work?

The Purair PCR Workstation uses HEPA-filtered vertical airflow to create an ISO Class 5 clean work zone. Room air passes through a pre-filter and HEPA filter before flowing downward across the work surface, helping protect samples and reagents from airborne contamination.

Unlike the PCR Workstation, the Purair UV Dead Air Box does not use a fan or filtered airflow. It provides a circulation-free, still-air enclosure with built-in UV contamination control.

If You Need...

HEPA-filtered clean air
Purair PCR Workstation
HEPA-filtered vertical airflow supplies clean air to the work zone.
An ISO Class 5 work zone
Purair PCR Workstation
The PCR work zone is specified as ISO Class 5.
Vertical clean airflow
Purair PCR Workstation
Active vertical airflow at a published initial average of 0.30 m/s (60 fpm).
A circulation-free enclosure
Purair UV Dead Air Box
No fan or forced airflow is used.
No airflow across sensitive work
Purair UV Dead Air Box
The enclosed work zone is intentionally still air.
Built-in UV contamination control
Both Products
Both incorporate 254 nm UV, timer and UV safety features.
Personnel protection from biohazards
Neither Product
Evaluate an appropriate biological safety cabinet.
Personnel protection from chemical vapors
Neither Product
Evaluate an appropriate chemical fume hood.

When Is a PCR Workstation a Good Choice?

When HEPA-Filtered Air Matters

Continuously filters incoming room air before supplying clean air to the work zone.

Provides an ISO Class 5 clean work zone for contamination-sensitive PCR and sample preparation.

Uses controlled downward airflow to help protect samples and reagents from airborne contamination.

When Is a Dead Air Box a Good Choice?

When Still Air Matters

Provides an enclosed, still-air environment without a fan or forced airflow.

Designed for workflows where filtered airflow is not required or desired.

Includes built-in 254 nm germicidal UV for supplemental surface decontamination between procedures.

Compare PCR Workstations & UV Dead Air Boxes

HEPA & ULPA Filtration

Air Science PCR Workstations and Dead Air Boxes help reduce contamination during PCR, DNA/RNA and sample preparation. Purair PCR provides HEPA-filtered vertical airflow for an ISO Class 5 clean work zone, while Purair Dead Air Boxes offer a circulation-free enclosure with built-in UV decontamination. Both are designed for non-hazardous applications where sample protection is essential.

UV Decontamination

Both Purair PCR Workstations and UV Dead Air Boxes feature built-in germicidal UV lamps to support surface decontamination between procedures. UV treatment is a supplemental contamination-control measure and should not replace routine cleaning, good laboratory practices or established decontamination protocols.

Primary Type
PCR clean bench
Dead air box / UV enclosure
Airflow
Vertical laminar airflow
Circulation-free, still air
Filtration
HEPA standard; ULPA optional
No filtration
Cleanliness
ISO Class 5 clean work zone
No ISO Class 5 filtered-air claim
UV Decontamination
Standard
Standard
Primary Approach
Filtered airflow + UV
Enclosed still air + UV
Typical Fit
PCR and sample preparation requiring filtered clean air
PCR and DNA/RNA preparation suited to still air
Personnel Protection
No
No
Available Widths
24", 36", 48"
24", 36", 48"

Explore Our PCR Workstations & Dead Air Boxes

PCR Workstations & UV Dead Air Boxes

Made in America by the New America

Frequently Asked Questions

Below are answers to our frequently asked questions and links to additional helpful information.

Purair PCR Workstations provide HEPA-filtered vertical airflow to create an ISO Class 5 clean work zone. Purair UV Dead Air Boxes use a circulation-free, still-air enclosure with built-in UV contamination control.

Neither is universally better. Choose a PCR Workstation when HEPA-filtered clean airflow and ISO Class 5 conditions are needed. Choose a UV Dead Air Box when a circulation-free enclosure is appropriate and filtered airflow is not required.

Yes. Purair PCR Workstations provide an ISO Class 5 clean work zone through HEPA-filtered vertical airflow. UV Dead Air Boxes do not have an ISO Class 5 filtered-air work-zone classification.

Purair PCR Workstations feature standard HEPA filtration rated at 99.97% efficiency at 0.3 microns, with optional ULPA filtration. Purair UV Dead Air Boxes do not use HEPA filtration or forced airflow.

Yes. Both platforms include built-in 254 nm germicidal UV lamps with timers and safety features to support surface decontamination between procedures.

No. UV is a supplemental contamination-control feature and should not replace routine cleaning, good laboratory practices or established decontamination procedures.

Yes. Both are designed for suitable non-hazardous, contamination-sensitive applications, including PCR preparation and DNA/RNA manipulation where personnel protection is not required.

No. Both products are intended for sample and product protection, not personnel containment. Applications involving biohazardous materials or hazardous chemical vapors require appropriate containment equipment.

No. A laminar flow cabinet uses controlled filtered airflow. A UV Dead Air Box intentionally operates without forced airflow.

Both Purair PCR Workstations and Purair UV Dead Air Boxes are available in 24″, 36″ and 48″ widths.

Yes. Both product families support benchtop use, with stand or mobile-cart configurations available depending on the model.

Start with your contamination-control requirements. Choose Purair PCR for HEPA-filtered ISO Class 5 airflow or a Purair UV Dead Air Box for a circulation-free enclosure. Then consider equipment dimensions, workflow and whether personnel protection is required.

Which PCR Contamination-Control Solution Is Right for You?

Air Science can help you determine whether a HEPA-filtered PCR Workstation or a circulation-free UV Dead Air Box is the right choice for your application, based on your airflow, filtration and contamination-control requirements.